Calibration System Basic, Up to 3 reference gauges, DAkks certified PN:PSA30105

$32,675.00

$27,675.00

Pfeiffer Vacuum Calibration System Basic, Up to 3 reference gauges, DAkks certified

Brochure: Calibration System Basic

Check out our new video of the Pfeiffer Vacuum Gauge Calibration System

 Features:

  • For static and dynamic calibration
  • Calibration range 1013 - 10-4 hPa
  • Easy to use
  • Chamber complies to ISO/TS 3567
  • Up to 3 reference gauges for medium accuracy
  • DAkkS certified reference gauges
  • Flexible
  • Customized systems available
Accuracy Depending on the used reference gauges
Cable length 3 m
Connection for gauge 2x DN 16 ISO-KF, 2x DN 25 ISO-KF, 2x DN 40 ISO-KF
Gas flow max. 5 · 10-6 - 1 · 103 hPa l/s
Gas inlet DN 16 ISO-KF
Material Stainless steel, gas bead blasted
Measurement range max. 1,000 · 100 hPa | 750 · 100Torr | 1,000 · 100 mbar
Measurement range min. 1 · 10-4 hPa | 7.5 · 10-5 Torr | 1 · 10-4mbar
Mode of operation Manual
Power consumption 230 VA
Pumping speed backing pump at 50 Hz 0.9 m3/h
Pumping speed for N2 67 l/s
Ultimate pressure <1 · 10-7 hPa | <7.5 · 10-8 Torr | <1 · 10-7 mbar
Voltage: Range 115 V, 50/60 Hz; 230 V, 50/60 Hz
Weight 27 kg | 59.52 lb

 Applications

  • Freeze-drying in the pharmaceutical and food industry
  • Coating systems for optics, ophthalmics, decorative coatings, diffusion barrier layers for PET bottles and packaging films, wear protection coatings for tools, etc., incorporating the use of multiple measuring instruments. 
  • Final vacuum measurement in pump production and service workshops
  • Process monitoring and process control for determining the end of the process, in order to produce consistent quality and minimize rejection rates. 

Calibration methods

Calibration is based on a comparison of the measurement values of the gauge to be calibrated and the reference gauge according to ISO/TS 3567. The gauge to be calibrated and the calibrated reference gauges are connected to a standardized vacuum chamber (according to ISO/TS 3567, form b) and are exposed to the same vacuum. 

There are two basic methods of calibration – static and dynamic calibration. With the calibration systems supplied by Pfeiffer Vacuum, transmitters can be calibrated statically with a measuring range of > 1 hPa as well as dynamically with a measuring range of < 1 hPa.

During calibration with 3 reference gauges, the following calibrating accuracy is obtained:

  • 0.2% accuracy at 1,013 hPa
  • 0.8% accuracy at 2 hPa 
  • 0.2% accuracy at 0.1 hPa
  • 10% accuracy at 10-4 hPa

Calibration with 3 reference gauges obtains an accuracy of 0.2 – 0.8% in the pressure range from atmospheric pressure (1,013 hPa) to 0.1 hPa.